Here’s how to do a simple comparison calibration of a probe-style?RTD?at one temperature point using?temperature sensor calibration equipment, a?dry-block calibrator?in this case.? Fully insert the probe RTD into one of the dry-block insert holes and connect the cable to an RTD temperature readout.? Ramp the dry block to 100 oC (212 oF).? When the dry-block display shows a stable 100 oC, record the value of the RTD temperature readout.? You now have one data point to compare the RTD temperature on the readout to the dry-block calibrator which is being used as a stable measurement reference standard.? A typical sensor calibration would be done at several temperature points that span the working range of the sensor when used in a production environment
Comparison calibration using a dry-block calibrator
For a higher accuracy calibration, a reference?PRT?connected to a thermometer readout could be inserted into another of the dry-block insert holes.? In this case, the PRT which has a higher accuracy than the internal display of the dry-block calibrator is used as the reference standard and compared to the reading of the RTD temperature readout.?
Comparison calibration using a reference PRT with dry-block
Please read?Selecting a dry-block calibrator?for help in choosing the right dry-block calibrator. ?