5500E Multi-Product Calibrator

  • No image available
已停產
本產品已停產或重新分類

產品概述: 5500E Multi-Product Calibrator

All the functions you need


The 5500E has a wide range of outputs available from a single calibration source. These include dc and ac voltages up to 1020 V, dc and ac currents up to 550 A (using the 5500A/COIL, 50 turn current coil accessory) and variable resistances up to 330 MΩ. The 5500E also provides continuously variable capacitance values up to 1.1 mF. For temperature calibration applications, thermocouples and RTDs are simulated and thermocouple signals can be measured. It also generates digitally synthesized extended bandwidth sine, truncated sine, triangle and square waveforms. The 5500E gives you broad workload coverage in one economically priced package.

Easy-to-use operation


While the Fluke 5500E calibrator helps you do more work, it also makes it easier to get work done. Its intuitive design makes operating it as natural as turning on a light switch, reducing references to the manual. For most tasks, your hand moves from the left to right, keeping you from having to make long, illogical or uncomfortable movements. Most functions require minimal keystrokes. The 5500E is also easy on your budget because no additional fixtures are needed and you can use it with conventional test leads.

Automate calibration for documented, consistent efficiency


With quality management standards like ISO 9000, there is a lot more to calibration than just making measurements. You also have documentation, control and reporting requirements to meet. Optional Microsoft Windows?-based Fluke metrology software simplifies the documentation of your procedures, adequacy and traceability. It also collects and reports calibration results information and helps you to consistently, quickly and efficiently calibrate a wide variety of instruments. With it your entire calibration process—from creating and executing procedures through results data collection and reporting—can be automated. Automated metrology is done by using a suite of applications called MET/CAL? Plus Calibration Management Software. Used with the MET/BASE-7 metrology database, the 5500/CAL-L license provides RS-232 instrument control and operates with the Fluke 5500A, 5520A, and 5800A calibrators in addition to the 5500E. If you need IEEE instrument control or want to automate the calibration of additional instruments, you can step up to GIPB based instrument control with the MET/CAL-L license.

產品規格: 5500E Multi-Product Calibrator

Technical Specifications
Voltage DCRange:?0 to ±1020V
Best 1 Year Specification:???? ±50 ppm of setting
Current DCRange:?? 0 to ±11A
Best 1 Year Specification:???? ±0.01% of setting
ResistanceRange:??? 0 to 3.29.999 MΩ
Best 1 Year Specification:???? ±0.009% of setting
Voltage ACRange:?? 1 mV to 1020V 10 Hz to 500 kHz, Sine
Best 1 Year Specification:???? ±0.03% of setting
Current ACRange:??? 29 mA to 11A 10 Hz to 10 kHz, Sine
Best 1 Year Specification:???? ±0.06% of setting
CapacitanceRange:?? 0.33 nF to 1.1 mF
Best 1 Year Specification:? ?? ±0.25% of setting
Thermocouple Source and MeasureRange:??-250°C to +2316°C (11 types)
Best 1 Year Specification:???? ± 0.14°C of setting
RTD SourceRange:???-200°C to +630°C (8 types)
Best 1 Year Specification:???? ±0.03°C of setting
FrequencyRange:??0.01 Hz to 2.0 MHz
Best 1 Year Specification:???? ±25 ppm of setting
General Specifications
Warm up TimeTwice the time since last warmed up, to a maximum of 30 minutes
Settling TimeLess than 5 seconds for all functions and ranges except as noted
Standard InterfacesIEEE-488 (GPIB), RS-232, 5725A (5500E only)
Temperature PerformanceOperating: 0°C to 50°C
Calibration: (tcal): 15°C to 35°C
Storage: -20°C to 70°C
Temperature CoefficientTemperature coefficient for temperatures outside tcal ± 5°C is 10% of the 90-day specification (or 1-year, as applicable) per °C
Relative HumidityOperating: < 80% to 30°C, < 70% to 40°C, < 40% to 50°C
Storage: < 95%, non-condensing
AltitudeOperating: 3,050m (10,000 ft) maximum
Non-operating: 12,200m (40,000 ft) maximum
SafetyDesigned to comply with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994; CAN/CSA-C22.2 No. 1010.1-92
Analog Low Isolation20V
EMCDesigned to comply with FCC Rules Part 15
Line PowerLine Voltage (selectable): 100V, 120V, 220V, 240V
Line Frequency: 47 to 63 Hz
Line Voltage Variation: : ± 10% about line voltage setting
Power Consumption300 VA
DimensionsHeight: 17.8 cm (7 in), standard rack increment,
????plus 1.5 cm (0.6 in) for feet on bottom of unit
Width: 43.2 cm (17 in), standard rack width
Depth: 47.3 cm (18.6 in) overall
Weight (without options)20 kg (44 lbs.)
Absolute Uncertainty Definition5500E, 5500A and 5520A uncertainty specifications include stability, temperature coefficient, linearity, line and load regulation, and the traceability of the external standards used for calibration. You do not need to add anything to determine the total uncertainty of your calibrator for the temperature range indicated.?
Specification Confidence Interval> 99%